Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Notes
Influence of pH for the Determination of Serum Albumin by a Dye-binding Method in the Presence of a Detergent
Yuji SUZUKI
著者情報
ジャーナル フリー

2008 年 24 巻 8 号 p. 1061-1064

詳細
抄録
In the dye-binding method, the absorbance increase caused by a protein error of a pH indicator is observed only in a restricted pH range. However, this pH range in the presence of a detergent has not yet been examined. Thus, the author investigated the pH (pHUL) where the absorbance increase becomes zero by a calculation based on the chemical equilibrium of a protein error of a pH indicator, and by experiments using four sulfonephthalein dyes. The pHUL value changed only with the detergent concentration, but did not change at all due to the dye, buffer solution or protein concentrations. Although the pHUL value was different according to the kind of dye used, it correlated well with the pKD values (dissociation constant) of BPB, BCG, BCP and BTB. The characteristics of pHUL in the reactions of the four dyes indicated good agreement with that obtained by a calculation.
著者関連情報
© 2008 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback