Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
A Miniaturized Stepwise Injection Spectrophotometric Analyzer
Anastasiia PETROVAAndrey BULATOVAndrey VISHNIKINLeonid MOSKVINRyoichi ISHIMATSUKoji NAKANOToshihiko IMATO
著者情報
ジャーナル フリー

2015 年 31 巻 6 号 p. 529-533

詳細
抄録
A novel micro-stepwise injection analyzer (μSWIA) has been developed for the automation and miniaturization of spectrophotometric analysis. The main unit of this device is a mixing chamber (MC) connected to the atmosphere. This part of the μSWIA provides rapid and effective homogenization of the reaction mixture components and completion of the reaction by means of gas bubbling. The μSWIA contained a rectangular labyrinth channel designed in way allowing one to eliminate bubbles by moving a solution from the MC to an optical channel. The light-emitting diode (LED) was used as a light emitter and the analytical signal was measured by a portable spectrophotometer. Fluid movement was attained via the use of a computer-controlled syringe pump. The μSWIA was successfully used for the spectrophotometric determination of cysteine in biologically active supplements and fodder by using 18-molybdo-2-phosphate heteropoly anion (18-MPA) as the reagent.
Fullsize Image
著者関連情報
© 2015 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top