Novel Device Accurately Measures Graft Resistance and Compliance to Ensure Quality of Coronary Artery Bypass
Kinichi NakataYukihiko OrimeKenji AkiyamaHayato KobaYoshiyuki SankaiMotomi Shiono
Author information
JOURNAL FREE ACCESS

2012 Volume 18 Issue 5 Pages 438-443

Details
Abstract

Introduction: The purpose of this study is to know the influence of coronary artery bypass grafting (CABG) on coronary circulation. In the present study, we evaluated CABG by using a novel flow analyzer that can calculate bypass graft resistance (Ra), resistance of the peripheral bed to which graft connects (Rp), the inertia of blood flow through the graft (L) and vascular wall compliance (C).
Methods: We performed off-pump CABG surgery on fifteen pigs assigned to the following groups (n = 5 each): normal CABG, competitive flow grafts and constrictive grafts.
Results: The wave pattern of 3 groups showed a clearly different form. In normal CABG and competitive flow group, we accepted a statistical difference in Rp and flow. In normal CABG and constrictive grafts. We accepted a statistical difference in Ra and flow.
Conclusion: We can know the relationship between CABG and coronary circulation by this device in detail. This device will be useful for evaluating graft performance during CABG.