Annals of Thoracic and Cardiovascular Surgery
Online ISSN : 2186-1005
Print ISSN : 1341-1098
ISSN-L : 1341-1098
Case Reports
Pure Cusp Tear of Trifecta Bioprosthesis 2 Years after Aortic Valve Replacement
Masaki HamamotoTaira KobayashiMasamichi OzawaKosuke Yoshimura
Author information
JOURNAL FREE ACCESS

2017 Volume 23 Issue 3 Pages 157-160

Details
Abstract

Trifecta is a stented bioprosthetic heart valve with a bovine pericardial sheet externally mounted on a titanium stent. This valve is applied only for aortic valve replacement (AVR), providing excellent hemodynamics and extremely low incidence of structural valve deterioration (SVD). A 76-year-old woman presented with dyspnea on effort 24 months after AVR with a 21-mm Trifecta valve. Echocardiography revealed severe aortic regurgitation with prolapse of a cusp of Trifecta valve, which suggested that she developed acute heart failure due to early SVD. In the operation, Trifecta valve had a cusp tear near the commissure with circumferential fibrous pannus ingrowth only at the inflow side. There was neither calcification nor infection. The Trifecta valve was successfully replaced with a new porcine bioprosthesis.

Content from these authors
© 2017 The Editorial Committee of Annals of Thoracic and Cardiovascular Surgery

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/