Annals of Vascular Diseases
Online ISSN : 1881-6428
Print ISSN : 1881-641X
ISSN-L : 1881-641X
Case Report
Surgical Treatment of Complications Associated with the Angio-Seal Vascular Closure Device: Report of Three Cases
Yasuyuki BitoMasayuki SakakiKazushige InoueYoshiteru YoshiokaHiroki Mizoguchi
著者情報
ジャーナル フリー

2010 年 3 巻 2 号 p. 144-147

詳細
抄録
The Angio-Seal arterial closure device consists of several bioabsorbable components and is used for hemostasis of arterial puncture sites. We report 3 cases of hemorrhagic and ischemic complications related to Angio-Seal use. Two cases were treated successfully by surgical removal of the device. In the third case surgical removal of the device failed and additional intervention was necessary. The unique structure of the Angio-Seal and the most likely cause of failure should be considered when treating device-related complications.
著者関連情報
© 2010 Japanese College of Angiology, The Japanese Society for Vascular Surgery, Japanese Society of Phlebology

この記事はクリエイティブ・コモンズ [表示 - 非営利 - 継承 4.0 国際]ライセンスの下に提供されています。
https://creativecommons.org/licenses/by-nc-sa/4.0/deed.ja