Agricultural and Biological Chemistry
Online ISSN : 1881-1280
Print ISSN : 0002-1369
ISSN-L : 0002-1369
Flow Injection Analysis of Lactose in Milk Using a Chemically Modified Lactose Electrode
Kiyoshi MATSUMOTOOsamu HAMADAHiroyuki UKEDAYutaka OSAJIMA
著者情報
ジャーナル フリー

1985 年 49 巻 7 号 p. 2131-2135

詳細
抄録
β-Galactosidase and glucose oxidase were immobilized with bovine serum albumin using glutaraldehyde on to a glassy carbon electrode silanized with 3-aminopropyltriethoxysilane. The laboratory-constructed lactose electrode was used for flow injection analysis to determine the lactose content in milk. Electrochemical interference could be detected by a non-enzymatic electrode (W2) and the current was subtracted from that of the enzymatic electrode (W1), providing an accurate measurement of the hydrogen peroxide that was enzymatically generated. The peak current was linearly related to the lactose concentration in the range 10-4-1.5 × 10-3M (original concentration) and 40 samples/hr could be analyzed. The relative standard deviation for 10 assays was less than 2%. The proposed method was compared with the chloramine T method and the values determined by both methods were in good agreement.
著者関連情報

この記事は最新の被引用情報を取得できません。

© Japan Society for Bioscience, Biotechnology, and Agrochemistry
前の記事 次の記事
feedback
Top