High-Performance Liquid Chromatographic Assay for the Determination of Nilotinib in Human Plasma
Misato YukiYuji YamakawaTakashi UchidaTakeru NambuTatsuya KawaguchiAkinobu HamadaHideyuki Saito
著者情報
ジャーナル フリー

2011 年 34 巻 7 号 p. 1126-1128

詳細
抄録
A precise and convenient high-performance liquid chromatography (HPLC) method has been estab-lished to assay nilotinib in human plasma. Chromatographic separation of nilotinib was performed on a LiChrosphere®100 RP-18(e) column (250 mm×4.0 mm, 5 μm) using a mixture of acetonitrile and 0.01 M phosphate buffer (pH 3.0) (42 : 58, v/v) under isocratic conditions at a flow rate of 1.0 ml/min with ultraviolet (UV) detection at 266 nm. The calibration curve showed linearity at concentrations between 250 ng/ml and 5000 ng/ml (r2>0.999). The mean±S.D. absolute recovery of nilotinib from plasma was 99.2±3.3%. The coefficients of variation of both intra- and inter-day precision were below 9.1%. These results indicate that this new HPLC-based quantification may be useful for therapeutic drug monitoring of nilotinib to help manage treatment in patients with chronic myeloid leukemia in clinical practice.