指標物質を用いる校正方法のための反応ガスクロマトグラフの試作と評価
渡邉 卓朗加藤 健次松本 信洋前田 恒昭
著者情報
ジャーナル フリー

2003 年 52 巻 4 号 p. 265-268

詳細
抄録

In order to realize a technique that serves as a calibration method using an index compound, reaction gas chromatography was examined. An index compound represents a group which includes many compounds; compounds in this group can be calibrated using the index compound. The gas chromatograph/micro reactor/mass spectrometer (GC/MR/MS) consists of a capillary gas chromatograph/mass spectrometer and a micro reactor. By using the MR, alkenes and alkynes were transformed to alkanes having the same carbon skeleton. Ethylene in a cylinder was determined using ethane as an index compound. The results showed good agreement with a reference value. The hydrodechlorination of chlorobenzenes was examined. Chlorobenzene in a solution was determined by using benzene as an index compound. The experimental results and reference concentration showed a good agreement. The efficacy of this technique was demonstrated with these results.