分析化学
Print ISSN : 0525-1931
定電位解法の研究(第6報)高感度自働定電位装置
石橋 雅義藤永 太一郎
著者情報
ジャーナル フリー

1954 年 3 巻 2 号 p. 96-98

詳細
抄録

In the preceeding paper (this journal 2, 345 (1953)) these authors described an apparatus of regulating circuit consisting of galvanometer switch-amplifier-electromagnetic relay-reversible motor, which used direct current as an electrolysis source. In the present paper, a new sensitive potentiostat employing galvano-meter switch-A. C. amplifier-reversible motor is described and also some results of using it are given. This instrument is completely line-operated from 110 volts alternating current source except one 1.5 volts dry battery as a supplement for cathode potential; no regulation is necessary before use and the potential of working electrode is kept automatically definite during the electrolysis. The working capacity for electrolysis is 30V and 3A, and the sensitivity to the change in electrolysis is about ±7 mV.

著者関連情報
© The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top