Circulation Journal
Online ISSN : 1347-4820
Print ISSN : 1346-9843
ISSN-L : 1346-9843
Clinical Investigation
Angiographic and Clinical Outcomes of a Pharmacokinetic Study of Sirolimus-Eluting Stents
Lesson From Restenosis Cases
Masato NakamuraMasamichi WadaHisao HaraKen KozumaYoritaka OtsukaShunichi Miyazaki
Author information
JOURNAL FREE ACCESS

2005 Volume 69 Issue 10 Pages 1196-1201

Details
Abstract
Background A pharmacokinetic (PK) study was conducted to evaluate sirolimus-eluting stents (SES) in Japanese people, representing the first clinical trial of the use of drug-eluting stents in Japan. Methods and Results The PK study was conducted in 20 patients with 30 lesions treated with sirolimus-coated BX Velocity stents. All lesions were treated with a single SES (3×18 mm). Angiographic follow-up was performed at 8 months after SES implantation, and the clinical outcomes were evaluated at 1 year in all cases. All procedures were successful, and all patients were discharged without any adverse cardiac events. The total restenosis rate was 10% (3 lesions) and target vessel revascularization was performed in those 3 cases (15%). Restenoses occurred at the proximal and distal stent margins. Intravascular ultrasound examination of restenosis cases revealed abundant plaque burden at the stent edges even though the luminal area was preserved. Conclusions The sirolimus-eluting BX Velocity stent is safe and useful for Japanese patients with coronary artery disease. However, restenosis at proximal stent edge seems to be a problem. (Circ J 2005; 69: 1196 - 1201)
Content from these authors
© 2005 THE JAPANESE CIRCULATION SOCIETY
Previous article Next article