Damage to Polymer of Paclitaxel-Eluting Stent
Hideki KitaharaYoshio KobayashiMasashi YamaguchiKenji OhkuboTakashi NakayamaNakabumi KurodaIssei Komuro
Author information
JOURNAL FREE ACCESS

2008 Volume 72 Issue 11 Pages 1907-1908

Details
Abstract
The polymer of paclitaxel-eluting stents (PES) plays an important role in controlling the release of paclitaxel. Damage to the polymer of a PES that is used in a patient has not been demonstrated, although in-vitro studies report disruption of it. The present case report describes damage to a PES as delivery through a calcified coronary artery was being attempted. (Circ J 2008; 72: 1907 - 1908)