Circulation Journal
Online ISSN : 1347-4820
Print ISSN : 1346-9843
ISSN-L : 1346-9843
Arrhythmia/Electrophysiology
Relationship Between Oral Amiodarone and Inappropriate Therapy From an Implantable Cardioverter Defibrillator
Takayuki NagaiKazuhiro SatomiTakashi NodaHideo OkamuraYuko YamadaWataru ShimizuKazuhiro SuyamaNaohiko AiharaShiro KamakuraTakashi Kurita
Author information
JOURNAL FREE ACCESS

2010 Volume 74 Issue 7 Pages 1302-1307

Details
Abstract
Background: This study evaluated the efficacy of amiodarone for avoiding inappropriate therapies by implantable cardioverter defibrillators (ICDs). Methods and Results: A total of 232 patients with structural heart disease (58±13 years; 78% males) who underwent an initial ICD implantation were retrospectively investigated to compare baseline characteristics and event rates of inappropriate ICD therapy delivery between patients with oral amiodarone therapy (amiodarone group, n=116) and those without (non-amiodarone group, n=116). During a mean follow-up of 29±21 months, inappropriate therapies occurred less frequently in the amiodarone group than in the non-amiodarone group (12% vs 27%, P=0.0068). As a cause of inappropriate ICD therapy, only atrial fibrillation (AF) significantly differed between the groups (3% vs 12%, P=0.01). The results of multivariate logistic regression analysis showed that amiodarone therapy (odds ratio (OR) 0.38, 95% confidence interval (CI) 0.19-0.77, P=0.0073) and no history of spontaneous AF (OR 0.27, 95%CI 0.13-0.57, P=0.0007) were independent predictors of a lower risk of inappropriate ICD therapy. Conclusions: In the present group of ICD patients with structural heart disease, inappropriate therapy delivery occurred predominantly in those with spontaneous AF and/or without amiodarone.  (Circ J 2010; 74: 1302 - 1307)
Content from these authors
© 2010 THE JAPANESE CIRCULATION SOCIETY
Previous article Next article
feedback
Top