Optimization of Coronary Attenuation in Coronary Computed Tomography Angiography Using Diluted Contrast Material
Naoto KawaguchiAkira KurataTeruhito KidoYoshiko NishiyamaTomoyuki KidoMasao MiyagawaAkiyoshi OgimotoTeruhito Mochizuki
Author information
JOURNAL FREE ACCESS

2014 Volume 78 Issue 3 Pages 662-670

Details
Abstract

Background: The purpose of this study was to evaluate a personalized protocol with diluted contrast material (CM) for coronary computed tomography angiography (CTA). Methods and Results: One hundred patients with suspected coronary artery disease underwent retrospective electrocardiogram-gated coronary CTA on a 256-slice multidetector-row CT scanner. In the diluted CM protocol (n=50), the optimal scan timing and CM dilution rate were determined by the timing bolus scan, with 20% CM dilution (5ml/s during 10s) being considered suitable to achieve the target arterial attenuation of 350 Hounsfield units (HU). In the body weight (BW)-adjusted protocol (n=50, 222mg iodine/kg), only the optimal scan timing was determined by the timing bolus scan. The injection rate and volume in the timing bolus scan and real scan were identical between the 2 protocols. We compared the means and variations in coronary attenuation between the 2 protocols. Coronary attenuation (mean±SD) in the diluted CM and BW-adjusted protocols was 346.1±23.9 HU and 298.8±45.2 HU, respectively. The diluted CM protocol provided significantly higher coronary attenuation and lower variance than did the BW-adjusted protocol (P<0.05, in each). Conclusions: The diluted CM protocol facilitates more uniform attenuation on coronary CTA in comparison with the BW-adjusted protocol.  (Circ J 2014; 78: 662–670)

Content from these authors
© 2014 THE JAPANESE CIRCULATION SOCIETY
Previous article Next article
feedback
Top