Impact of Implantation Technique and Plaque Morphology on Strut Embedment and Scaffold Expansion of Polylactide Bioresorbable Scaffold – Insights From ABSORB Japan Trial –
Yohei SotomiYoshinobu OnumaJouke DijkstraJeroen EggermontShengnan LiuErhan TenekeciogluYaping ZengTaku AsanoRobbert J. de WinterJeffrey J. PopmaKen KozumaKengo TanabePatrick W. SerruysTakeshi Kimura
Author information
JOURNAL FREE ACCESS FULL-TEXT HTML

2016 Volume 80 Issue 11 Pages 2317-2326

Details
Abstract

Background:The optimal implantation technique for the bioresorbable scaffold (Absorb, Abbott Vascular) is still a matter of debate. The purpose of the present study was to evaluate the effect of implantation technique on strut embedment and scaffold expansion.

Methods and Results:Strut embedment depth and scaffold expansion index assessed by optical coherence tomography (OCT) (minimum scaffold area/reference vessel area) were evaluated in the ABSORB Japan trial (OCT subgroup: 87 lesions) with respect to implantation technique using either quantitative coronary angiography (QCA) or OCT. Strut embedment was assessed at the strut level (n=667), while scaffold expansion was assessed at the lesion level (n=81). The mean embedment depth was 63±59 µm. Balloon sizing and inflation pressure had no direct effect on strut embedment. Plaque morphology affected strut embedment [nonatherosclerotic (58.9±54.3 µm), fibroatheroma (73.3±59.6 µm), fibrous plaque (59.7±51.1 µm), and fibrocalcific plaque (–3.1±61.6 µm, negative value means malapposition), P <0.001]. The balloon-artery ratio positively correlated with the expansion index. This relationship was stronger when the OCT-derived reference vessel diameter (RVD) was used as a reference for balloon selection rather than the QCA-derived one [predilatation (Pearson correlation r: QCA: 0.167 vs. OCT: 0.552), postdilatation (QCA: 0.316 vs. OCT: 0.717)].

Conclusions:Underlying plaque morphology influenced strut embedment, whereas implantation technique had no direct effect on it. Optimal balloon sizing based on OCT-derived RVD might be recommended. However, the safety of such a strategy should be investigated in a prospective trial. (Circ J 2016; 80: 2317–2326)