Circulation Journal
Online ISSN : 1347-4820
Print ISSN : 1346-9843
ISSN-L : 1346-9843
Valvular Heart Disease
Transcatheter Aortic Valve Replacement With Different Valve Types in Elliptic Aortic Annuli
Yoshio MaenoYigal AbramowitzSung-Han YoonHasan JilaihawiSharma RaulSharjeel IsrarMasaki MiyasakaHiroyuki KawamoriYoshio KazunoTanya RamiNobuyuki TakahashiGeeteshwar MangatMohammad KashifTarun ChakravartyMamoo NakamuraWen ChengRaj R. Makkar
Author information
JOURNAL FREE ACCESS FULL-TEXT HTML

2017 Volume 81 Issue 7 Pages 1036-1042

Details
Abstract

Background:The aim of this study was to determine the influence of an elliptic annulus on acute device success rates following self-expanding (SE) transcatheter aortic valve replacement (TAVR) vs. balloon-expandable (BE) TAVR.

Methods and Results:Outcomes were assessed using Valve Academic Research Consortium-2 definitions. Aortic annulus ratio (AAR) was measured as short axis diameter/long axis diameter. Mean AAR was 0.81±0.06. Patients were therefore divided into 2 groups: AAR <0.82 and AAR ≥0.82. For circular annuli (AAR ≥0.82; 363 patients), high device success rates were achieved in both valve groups (SE valve, 90.5% vs. BE valve, 95.0%, P=0.14). Conversely, for AAR <0.82 (374 patients), SE valves had lower device success rates than BE valves (82.5% vs. 95.3%, P=0.002). For elliptic annuli, SE-TAVR was an independent predictor of unsuccessful device implantation (OR, 6.34, P<0.001). Nonetheless, increased oversizing of SE valves for elliptic annuli was associated with an exponential rise in device success (threshold ≥17.5%; area under the curve, 0.83) but not for BE-TAVR. Furthermore, optimally oversized SE valves and BE valves had a similarly high device success for elliptic annuli (SE valve, 96.2% vs. BE valve, 95.3%).

Conclusions:For circular annuli, similarly high device success was achieved for the 2 valve types. Conversely, for elliptic annuli, SE valves had a lower device success than BE valves. Device success following optimal oversizing of SE valves, however, was similar to that for BE valves.

Content from these authors
© 2017 THE JAPANESE CIRCULATION SOCIETY
Previous article Next article
feedback