2020 Volume 84 Issue 12 Pages 2312-2319
Background:We investigated the long-term clinical and hemodynamic outcomes after aortic valve replacement (AVR) with a 17-mm mechanical valve.
Methods and Results:Between January 2005 and December 2011, 80 patients with aortic stenosis underwent AVR with the 17-mm St. Jude Medical Regent prosthetic valve. Echocardiography was performed preoperatively, at discharge, and at follow-up, which was performed at least 2 years postoperatively (median interval, 7.3 years). Prosthesis-patient mismatch (PPM) was defined as an indexed effective orifice area <0.85 cm2/m2at discharge and occurred in 25 patients (31%). The median follow-up period was 8.7 years (100% complete). Overall in-hospital mortality was 2.5% (2 patients) with 27 late deaths (34%). The 5- and 10-year survival rates were 78.7% and 63.0%, respectively. Peripheral arterial disease and concomitant mitral valve repair were independent predictors of late mortality. The 5- and 10-year freedom from major adverse valve-related events (MAVRE) rates were 91.6% and 83.5%, respectively. PPM at discharge did not affect long-term survival, freedom from MAVRE, or freedom from heart failure. Echocardiographic data at follow-up revealed a significant reduction in the mean left ventricular mass index (LVMI). LVMI reduction observed at follow-up was similar between patients with and without PPM.
Conclusions:AVR with the 17-mm mechanical prosthesis had acceptable long-term clinical and hemodynamic outcomes. Significant reduction in LVMI was observed regardless of PPM.