Chemical and Pharmaceutical Bulletin
Online ISSN : 1347-5223
Print ISSN : 0009-2363
ISSN-L : 0009-2363
Regular Articles
Quantification of Terbutaline in Pharmaceutical Formulation and Human Serum by Adsorptive Stripping Voltammetry at a Glassy Carbon Electrode
Amr Mohamed Beltagi, Hana Salah El-Desoky, Mohamed Metwalli Ghoneim
著者情報
ジャーナル フリー

2007 年 55 巻 7 号 p. 1018-1023

詳細
抄録
The electrochemical oxidative behavior of terbutaline at the glassy carbon electrode was studied in a series of the Britton-Robinson buffer of pH 2—11. Cyclic and square-wave voltammograms of terbutaline at the pH values ≤9 exhibited a single irreversible anodic peak. A fully validated, simple, sensitive and precise square-wave adsorptive anodic stripping voltammetric procedure was described for the determination of terbutaline in bulk form, tablets and human serum. For 5×10−8 M bulk terbutaline a mean recovery of 98.78±0.94% (n=5) was achieved following its preconcentration by adsorptive accumulation onto the glassy carbon electrode at a +0.15 V (vs. Ag/AgCl/KCls) for 180 s. Limit of detection of 6×10−9 M and limit of quantitation of 2×10−8 M terbutaline were achieved in the bulk form or its formulations (Bricanyl® tablets). The described square-wave adsorption anodic stripping voltammetric procedure was successfully applied for the determination of terbutaline in human serum following medium exchange. Limit of detection of 1.41×10−8 M (3.173 ng ml−1) and limit of quantitation of 4.70×10−8 M (10.575 ng ml−1) were achieved in human serum with a mean recovery of 98.11±1.13%.
著者関連情報
© 2007 The Pharmaceutical Society of Japan
前の記事 次の記事
feedback
Top