Note
Effect of Noise Filtering in an Ultraviolet Detector of UHPLC on the Statistical Reliability of Chemometrically Evaluated Repeatability
Akira Kotani Koichi MachidaRyo WatanabeHideki Hakamata
Author information
JOURNAL OPEN ACCESS FULL-TEXT HTML
Supplementary material

2024 Volume 72 Issue 9 Pages 800-803

Details
Abstract

A noise filter, which is usually attached to a detector for chromatography, was applied for the improvement of a signal-to-noise ratio (S/N) on a chromatogram. The objective of this paper is to elucidate the effect of noise filtering in an UV detector of ultra HPLC (UHPLC) on the statistical reliability of chemometrically evaluated repeatability by the function of mutual information (FUMI) theory. To examine the statistical reliability of chemometrically evaluated repeatability in the UHPLC system associated with noise filtering, the standard deviation (SD) values of the area in baseline fluctuations with peak region k (s(k)) were obtained from six chromatograms with noise filtering. Further, the average of s(k) values (σ̂) was calculated from the s(k) values (n = 6) to be alternatively applied as the population SD. All s(k)/σ̂ values were within the 95% confidence intervals (CIs) at the freedom degree of 50, indicating the chemometrically estimated relative SD (RSD) of a peak area and RSD by repeated measurements of at least 50 times had equivalent reliability.

Fullsize Image
Content from these authors
© 2024 Author(s)
Published by The Pharmaceutical Society of Japan

This article is licensed under a Creative Commons [Attribution-NonCommercial 4.0 International] license.
https://creativecommons.org/licenses/by-nc/4.0/