Chemical and Pharmaceutical Bulletin
Online ISSN : 1347-5223
Print ISSN : 0009-2363
ISSN-L : 0009-2363
Electrochemical Detector for High-Performance Liquid Chromatography. V. Application to Adsorption Chromatography
OSAMU HIROSHIMASATORU IKENOYAMASAHIKO OHMAEKIYOSHI KAWABE
Author information
Keywords: menaquinone
JOURNAL FREE ACCESS

1981 Volume 29 Issue 2 Pages 451-455

Details
Abstract
Electrochemical detection was applied to high-performance liquid chromatography (adsorption chromatography). The column eluate was mixed with methanol-ethanol mixture containing NaClO4 as a supporting electrolyte and analyzed with an electrochemical detector. We applied this method to the determination of tocopherol isomers (TP) in wheat germ oil and phylloquinone isomers (PK) in rat plasma. The limits of detection were 1 ng and 10 ng for TP and PK, respectively. The present method can be applied to the determination of small amounts of geometricalisomers or stereoisomers
Content from these authors
© The Pharmaceutical Society of Japan
Previous article Next article
feedback
Top