Effect of Thermocycling on Tensile Strength and Tear Resistance of Four Soft Denture Liners
Serra OGUZMustafa Murat MUTLUAYOrhan Murat DOGANBulent BEK
著者情報
キーワード: Liner, Strength, Thermocycling
ジャーナル フリー

2007 年 26 巻 2 号 p. 296-302

詳細
抄録
This study evaluated the effect of thermocycling on the tensile strength and tear resistance of four long-term soft denture liners. One light-activated (Astron Light, AL), two chemically activated (GC Reline Soft, GC; Silagum Comfort, SC), and one heat-cured (Molloplast-B, MLP) soft liner materials were tested. Dumbbell and trouser-leg specimen geometries were used for tensile strength and tear resistance tests, respectively. A total of 120 specimens were prepared. Test specimens for each material (n=5) were subjected to thermal cycling for 1,000 and 3,000 cycles between 5°C and 55°C in a thermocycler. Before thermocycling, AL gave the lowest tensile strength, while SC exhibited the highest tear resistance value among the materials tested (p<0.05). Thermal cycling significantly affected the tensile strength of AL as well as the tear resistance values of AL, MLP, and GC materials. This in vitro study revealed that the tensile strength and tear resistance values of the soft liner materials tested varied according to their chemical compositions.