Three-dimensional finite element analysis of posterior fiber-reinforced composite fixed partial denture Part 2: influence of fiber reinforcement on mesial and distal connectors
Nobuhisa AIDAAkikazu SHINYADaiichiro YOKOYAMALippo V.J. LASSILAHarunori GOMIPekka K. VALLITTUAkiyoshi SHINYA
Author information
JOURNAL FREE ACCESS

2011 Volume 30 Issue 1 Pages 29-37

Details
Abstract
The aim of this study was to evaluate the influence of connectors under two different loading conditions on displacement and stress distribution generated in isotropic hybrid composite fixed partial denture (C-FPD) and partially anisotropic fiber-reinforced hybrid composite fixed partial denture (FRC-FPD). To this end, two three-dimensional finite element (FE) models of three-unit FPD from mandibular second premolar to mandibular second molar —intended to replace the mandibular first molar— were developed. The two loading conditions employed were a vertical load of 629 N (applied to eight points on the occlusal surface) and a lateral load of 250 N (applied to three points of the pontic). The results suggested that the reinforcing fibers in FRC framework significantly improved the rigidity of the connectors against any twisting and bending moments induced by loading. Consequently, maximum principal stress and displacement generated in the connectors of FRC-FPD were significantly reduced because stresses generated by vertical and lateral loading were transferred to the reinforcing fibers.
Content from these authors
© 2011 The Japanese Society for Dental Materials and Devices