Dental Materials Journal
Online ISSN : 1881-1361
Print ISSN : 0287-4547
ISSN-L : 0287-4547
Original Paper
Assessment of occlusal fissure depth and sealant penetration using optical coherence tomography
Sachiko ITOYasushi SHIMADAAlireza SADRYukie NAKAJIMAMichiyo MIYASHINJunji TAGAMIYasunori SUMI
Author information
JOURNAL FREE ACCESS

2016 Volume 35 Issue 3 Pages 432-439

Details
Abstract

Swept source optical coherence tomography (SS-OCT) is an interferometric imaging technique. This study aimed to assess SS-OCT as a diagnostic tool for the evaluation of occlusal fissure depth and sealant penetration into fissure. Seventy investigation sites of occlusal fissures without visible evidence of caries were categorized into four groups (smooth, shallow, intermediate, and deep fissures). Categorization utilized two methods: 1) visually, using a computer screen, and 2) using SS-OCT images. After sealant placement, penetration was observed in SS-OCT. The results obtained from SS-OCT and visual inspections were compared with those of confocal laser scanning microscope (CLSM). The diagnostic power of SS-OCT was higher than that of visual inspection for fissure depth. Additionally, clear cross-sectional images of sealant penetration into fissures were observed with SS-OCT. SS-OCT can be used to evaluate fissure depth and monitor sealant penetration.

Content from these authors
© 2016 The Japanese Society for Dental Materials and Devices
Previous article Next article
feedback