Influence of magnetic susceptibility and volume on MRI artifacts produced by low magnetic susceptibility Zr-14Nb alloy and dental alloys
Yuka KAJIMAAtsushi TAKAICHIYusuke TSUTSUMITakao HANAWANoriyuki WAKABAYASHIAkira KAWASAKI
Author information
JOURNAL FREE ACCESS

2020 Volume 39 Issue 2 Pages 256-261

Details
Abstract

The artifact volume generated in magnetic resonance (MR) images was quantitatively evaluated to investigate how artifact behavior correlates to the magnetic susceptibility and volume of an implanted metal device. For this, a new low-magnetic-susceptibility Zr-14Nb alloy was compared with two conventional dental alloys, Ti-6Al-7Nb alloy and Co-Cr-Mo alloy, using spherical specimens of each alloy prepared with four different diameters. Then, MR images were recorded under fast spin echo and gradient echo conditions, from which the artifact volume was measured. The artifact volume decreased with the magnetic susceptibility, volume, and mass of the specimens, and significant linear correlations were observed. The artifact volume can be estimated by the equations presented here; nevertheless, further studies are necessary to interpret the influence of some important factors (e.g., imaging conditions, shape, and orientation) to predict the artifact volume more precisely.

Content from these authors
© 2020 The Japanese Society for Dental Materials and Devices
Previous article Next article