Dental Materials Journal
Online ISSN : 1881-1361
Print ISSN : 0287-4547
ISSN-L : 0287-4547
Original Paper
Cyclic fatigue resistance for six types of nickel titanium instruments at artificial canals with different angles and radii of curvature
Danlu CHIYuejiao ZHANGXinwei LINZhongchun TONG
Author information
JOURNAL FREE ACCESS

2021 Volume 40 Issue 5 Pages 1129-1135

Details
Abstract

This study evaluated the cyclic fatigue resistance for six types of 25# NiTi instruments. A traditional manufacturing instrument, an M wire instrument, a gold treatment instrument, a controlled memory (CM) wire instrument, a CM wire instrument with electrical discharge machining (EDM), and an R-phase heat treatment instrument, were operated in the different curved artificial canals. The fracture time (FT) and number of cycles to fracture (NCF) of the NiTi instruments were higher at 45° angles and double-curvature canals than at 60° angles. Except for the instruments with gold technology and EDM technology, others showed the longest FT and the highest NCF at an 8 mm radius of curvature. Morphological characteristics of cyclic fatigue were exhibited on the cross-section and lateral view of fracture fragments. The use of M-wire, R-phase wire, CM-wire, gold technology, EDM technology, and reciprocating movement were beneficial to enhance the cyclic fatigue resistance of NiTi files.

Content from these authors
© 2021 The Japanese Society for Dental Materials and Devices
Previous article Next article
feedback