Dental Materials Journal
Online ISSN : 1881-1361
Print ISSN : 0287-4547
ISSN-L : 0287-4547
Original Paper
Effect of preset torque setting on torque/force generation, shaping ability and surface changes of nickel titanium rotary instrument in different root canal curvature locations: An ex vivo study
Moe Sandar KYAWArata EBIHARAYoshiko IINOMyint THUKeiichiro MAKIShunsuke KIMURAPyae Hein HTUNTakashi OKIJI
著者情報
ジャーナル オープンアクセス

2024 年 43 巻 3 号 p. 329-337

詳細
抄録

The aim of this study was to evaluate how preset torque settings influence the torque, vertical force, and root canal-centering ability of ProGlider and ProTaper NEXT nickel-titanium rotary instruments in canals with different curvature locations. Based on micro-computed tomography, mesial roots of human mandibular molars (25°–40° curvature) were allocated to the apical curvature (apical 1–5 mm) or the middle curvature (apical 5–9 mm) groups, and mandibular incisors (curvature <5°) to the straight canal group. Each group was subjected to automated instrumentation and torque/force measurement with the preset torque of 1, 2.5, or 5 N•cm. Canal-centering ratios were determined with micro-computed tomography. Instrument fracture occurred only in the 2.5 and 5 N•cm groups in curved canals. The preset torque setting and curvature location did not influence canal shaping ability.

著者関連情報
© 2024 The Japanese Society for Dental Materials and Devices

This is an open access article under the CC BY license
https://creativecommons.org/licenses/by/4.0/
次の記事
feedback
Top