The mechanical effect of aligner’s thickness and material properties in invisible orthodontics: A quantitative finite element study
Waheed AHMADJing XIONGZeyang XIA
Author information
JOURNAL OPEN ACCESS

2025 Volume 44 Issue 4 Pages 375-385

Details
Abstract

Invisible orthodontics benefits significantly from aligner technology, yet optimizing material properties and thickness for diverse patient needs is a challenge. The aim of this study was to investigate the interaction between key material properties and thickness—and their collective influence on orthodontic treatment outcomes. A three-dimensional model of the tooth, periodontal ligament, and bone complex was constructed, with attachments centered on each tooth crown. Nine aligners, varying in thickness from 0.20 to 1.00 mm with material properties (Young’s modulus (E) from 0.01 to 3.50 GPa and Poisson’s ratio (ν) at 0.30), were analyzed. The study measured force and moment changes due to a 0.15 mm mesial-distal movement of the lateral incisor, validated by other teeth. Results show aligner thickness and E significantly affect force and moment, with increases promoting translational rather than tipping movements. These findings underscore precise aligner design’s role in enhancing orthodontic outcomes.

Content from these authors
© 2025 The Japanese Society for Dental Materials and Devices

This is an open access article under the CC BY license
https://creativecommons.org/licenses/by/4.0/