Dental Materials Journal
Online ISSN : 1881-1361
Print ISSN : 0287-4547
ISSN-L : 0287-4547

This article has now been updated. Please use the final version.

A new method to measure the mechanical properties of rotary endodontic Ni-Ti files: Measurement of 6-axis force/torque and automatic root canal preparation
Hyungwoo LEEIn-Bog LEE
Author information
JOURNAL OPEN ACCESS Advance online publication

Article ID: 2024-082

Details
Abstract

The purpose of this study was to develop a new instrument to measure the mechanical properties of rotary endodontic Ni-Ti files (ProTaper Gold F2, ProTaper Ultimate F2, and HyFlex EDM Onefile), and to evaluate the overall utility of the device. The instrument was capable of analyzing the 6-axis force/torque generated by the files during cyclic dynamic movement in a metal curved artificial root canal, and doing automatic cyclic dynamic filing in a resin root canal with a preset vertical force limit by adopting a negative feedback mechanism. By analyzing the 6-axis force/torque, we were able to estimate the position and contact points of the files in the curved root canal. ProTaper Gold showed the highest force/torque in all directions. HyFlex EDM had the highest hysteresis ratio, centering ratio value and NCF (number of cycles to fatigue fracture), while the lowest vertical force.

Content from these authors
© 2024 The Japanese Society for Dental Materials and Devices

This is an open access article under the CC BY license
https://creativecommons.org/licenses/by/4.0/