Dental Materials Journal
Online ISSN : 1881-1361
Print ISSN : 0287-4547
ISSN-L : 0287-4547

この記事には本公開記事があります。本公開記事を参照してください。
引用する場合も本公開記事を引用してください。

Effects of repetitive insertion/removal and occlusal load on the retentive force of rest plate–I bar clasps made by selective laser melting
Yuki UEKUBOYoshimitsu KATOKeita TOMONOMitsuo KATOJuro WADACHIShinji TAKEMOTOShuichiro YAMASHITA
著者情報
ジャーナル オープンアクセス 早期公開

論文ID: 2024-271

この記事には本公開記事があります。
詳細
抄録

The objective of this study was to investigate the effects of repetitive removal/insertion and a load equivalent to the occlusal force on the retentive force of rest plate–I bar (RPI) clasps fabricated by selective laser melting (SLM) on powder bed fusion. RPI clasp fabricated by SLM and casting were evaluated to the retentive force by repetitive removal/insertion (SLM-RI and Cast-RI), and the retentive force of RPI clasp by SLM was also evaluated by repetitive removal/insertion under load (SLM-L). SLM-RI group was kept superior retentive force during repetitive removal/insertion than Cast-RI group. SLM-L group showed the decrease of retentive force, and all specimens fractured during less than 7,000 repetitive removal/insertion. In SLM-L, the fracture of RPI clasp occurred at minor connector due to stress concentration. In conclusion, the RPI clasp by SLM demonstrated the superior retentive force and fitting for repetitive removal/insertion, yet improvements were necessary to withstand a load equivalent to occlusal force.

著者関連情報
© 2025 The Japanese Society for Dental Materials and Devices

This is an open access article under the CC BY license
https://creativecommons.org/licenses/by/4.0/
feedback
Top