IEICE Electronics Express
Online ISSN : 1349-2543
ISSN-L : 1349-2543
LETTER
Detection of tissue coagulation for microwave surgical devices
Naoyuki OgasawaraKazuyuki Saito
Author information
JOURNAL FREE ACCESS

2017 Volume 14 Issue 3 Pages 20161223

Details
Abstract

This paper describes the detection of tissue coagulation for a microwave surgical device. This method uses variations in the electrical properties of biological tissue before and after heating. To evaluate the proposed method, we made coagulated tissue and measured the impedance of the tissue using the microwave surgical device. The results of the measurements show that the impedance of the heated tissue changed significantly due to tissue coagulation. This results implies that coagulation of tissue can be detected by variation of electrical impedance.

Content from these authors
© 2017 by The Institute of Electronics, Information and Communication Engineers
Previous article Next article
feedback