Low-Power Low-Data-Loss Bio-Signal Acquisition System for Intelligent Electrocardiogram Detection
Liang-Hung WangWei-Zhong DongJian-Zhi ChenFa-Xiang WangMing-Hui Fan
著者情報
ジャーナル フリー 早期公開

論文ID: 14.20161142

この記事には本公開記事があります。
詳細
抄録

The research describes an intelligent electrocardiogram (ECG) acquisition system with low power and low data loss features. The system is constructed with ADS129X, MSP430X, CC254X Bluetooth low energy (BLE) with low power wireless communication protocol and smart phone for the ECG display. Two architectures, namely, with and without the microprocessor (MPU) and hinted handoff mechanism algorithm, are used to verify the data transmission loss between the BLE and the smart phone. The system not only employed low-power solution but also added full handshake and hinted handoff to achieve complete synchronization and accurate transmission reliability. The MPU method definitely enhanced the ECG packet loss rate and power saving rate to 0.63% and 49.78%, respectively.