A low power QRS detection processor with adaptive scaling of processing resolution
Zhijian ChenMenghan JiaJiahui LuoTaotao ZhuXiaoyan XiangJianyi Meng
著者情報
ジャーナル フリー 早期公開

論文ID: 14.20170882

この記事には本公開記事があります。
詳細
抄録

This work presents a power efficient QRS detection processor with self-adjustable processing resolution. As reducing the resolution of electrocardiogram (ECG) data in certain degree has little effect on the detection accuracy but helps cut down the calculation power, a total of eight processing resolutions are supported in the processor, which can bring different levels of dynamic power reduction. The processing resolution is scaled adaptively based on the QRS detection result to achieve optimized energy efficiency and guarantee acceptable detection performance. The processor is implemented in SMIC 40nm CMOS technology and has a total area of 5655um2. When tested on MIT-BIH arrhythmia database with 50MHz operating frequency and 0.9V voltage supply, its power consumption is 77.7uW, which is reduced by 27.4% compared to the original processor, while the area is only increased by 6.7%. And it also achieves relatively high detection result, with an average sensitivity of 98.63% and positive prediction of 98.86%.