IEICE Electronics Express
Online ISSN : 1349-2543
ISSN-L : 1349-2543

この記事には本公開記事があります。本公開記事を参照してください。
引用する場合も本公開記事を引用してください。

Study of sleeve inner diameter to suppress backward heating for microwave surgical energy device
Tsugumi NishidateKazuyuki Saito
著者情報
ジャーナル フリー 早期公開

論文ID: 20.20230462

この記事には本公開記事があります。
詳細
抄録

In modern surgery, microwave surgical energy devices are used for hemostasis and anastomosis. However, elongated heating region called backward heating distributes around the feeding line, and potentially causes undesirable heating to non-targeted, normal tissue. In order to suppress the backward heating, we considered use of a sleeve with microwave device. Large sleeve inner diameter is required to use the sleeve effectively. However, small sleeve inner diameter is preferable for use in surgery. In this study, numerical analyses were performed to investigate the optimum sleeve inner diameter. The results indicated over 4 mm inner diameter could suppress the backward heating.

著者関連情報
© 2023 by The Institute of Electronics, Information and Communication Engineers
feedback
Top