Endocrine Journal
Online ISSN : 1348-4540
Print ISSN : 0918-8959
ISSN-L : 0918-8959

This article has now been updated. Please use the final version.

Impact of the transition from radioimmunoassay (RIA) to chemiluminescent enzyme immunoassay (CLEIA) for the measurement of plasma aldosterone concentration (PAC) on the diagnosis of primary aldosteronism (PA) via retrospective analyses in Okinawa, Japan
Ken-ichiro HonmaYoshiro NakayamaAtsuko TamakiMoriyuki UeharaTaiki TeruyaTakamitsu YabikuYohei IshikiKen YonahaRei ChinenTsugumi UemaShiki OkamotoHiroaki Masuzaki
Author information
JOURNAL OPEN ACCESS Advance online publication
Supplementary material

Article ID: EJ24-0227

Details
Abstract

In Japan, the traditional method for measuring plasma aldosterone concentration (PAC) was radioimmunoassay (RIA), which had several challenges, including poor traceability of certified reference materials and reduced detection sensitivity at low concentrations. To overcome these issues, a chemiluminescent enzyme immunoassay (CLEIA) for PAC measurement was introduced in April 2021 and the Japan Endocrine Society published new guidelines for primary aldosteronism (PA). This study aimed to evaluate the impact of the transition from RIA to CLEIA for PAC measurement on PA diagnosis. Data from 190 patients admitted to the Second Department of Internal Medicine, University of the Ryukyus Hospital, between April 2012 and March 2021 were analyzed. Patients who were diagnosed with PA underwent adrenal venous sampling. The PAC measured by RIA (PAC(RIA)) was converted to the estimated PAC measured by CLEIA (ePAC(CLEIA)) using a conversion formula. The present study evaluated the discordance rates in diagnoses based on screening (SC), captopril challenge test (CCT), saline infusion test (SIT), and diagnosis of PA between results judged by PAC(RIA) according to the previous guidelines and those judged by ePAC(CLEIA) according to the new guidelines. The results revealed discordant diagnosis rates of 6.4% for SC and 10.1% for CCT, with no discordance for SIT. The discordant diagnosis rate for PA was 3.7%. Our study reveals the challenges in establishing appropriate diagnostic criteria for PA using PAC(CLEIA) and highlights the demand for further research on provisionally positive categories.

Content from these authors
© The Japan Endocrine Society

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/deed.en
feedback
Top