地質学雑誌
Online ISSN : 1349-9963
Print ISSN : 0016-7630
ISSN-L : 0016-7630
ノート
微小なビトリナイト粒子のための反射率測定装置の製作
坂口 有人向吉 秀樹
著者情報
ジャーナル フリー

2012 年 118 巻 4 号 p. 240-244

詳細
抄録

Vitrinite reflectance geothermometer was originally developed for use in oil exploration, but it has various applications; e.g., estimating the depth of burial in sedimentary basins, analysis of the thermal structure of accretionary complexes, and detecting evidence of frictional heat in fault rocks. However, use of the vitrinite reflectance geothermometer is not widespread. This paper describes a modified apparatus that was specifically designed to analyze small vitrinite particles in accretionary complexes and fault rocks. The original apparatus was designed to illuminate large coal samples over a wide field of view (>100 μm), and measured reflectance with a high-voltage photomultiplier. A wide field of illumination is required for petrographic identification of vitrinite; however, reflections from other bright minerals around the small vitrinite particles generate a high level of background noise. The apparatus described here has a dual illumination system: a homogeneous beam illuminates a wide field of view to allow identification of vitrinite, and a concentrated spot beam of only 1.6 μm diameter is used to measure reflectance. The spot illumination system measures reflectance with a semiconductor photodiode sensor. The avalanche-type photodiode has a sensor cooling system, provides high-resolution, low-noise reflectance measurements, and is simpler to use than the photomultiplier.

著者関連情報
© 2012 日本地質学会
前の記事 次の記事
feedback
Top