日本歯科理工学会学術講演会要旨集
50th General Session of the Japanese Society for Dental Materials and Devices in conjunction with International Dental Materials Congress 2007
セッションID: P-213
会議情報

ポスター発表
Permeability of the 405-nm diode laser light into the dentin
*明石 豪加藤 純二末森 豪中澤 妙衣子平井 義人
著者情報
会議録・要旨集 フリー

詳細
抄録

In this study, we measured the permeability of the 405-nm diode laser light into the dentin for clinical use such as dental hypersensitivity or pulp capping. Forty extracted bovine mandibular anterior teeth were used. Dentin permeability was about 30% at thickness of 0.5mm and 15% at thickness of 1mm and 5% at thickness of 2.0mm.There was a correlation between the permeability and the thickness of the dentin. This result suggested that the 405-nm diode laser may become one of the options of the treatment of dental hypersensitivity or the inflamed pulp.

著者関連情報