電気学会論文誌C(電子・情報・システム部門誌)
Online ISSN : 1348-8155
Print ISSN : 0385-4221
ISSN-L : 0385-4221
<医用電子・生体工学>
バイオセンサ用銀-塩化銀電極の電極電位—温度特性
田頭 功関 加奈
著者情報
ジャーナル フリー

2007 年 127 巻 7 号 p. 990-994

詳細
抄録

The aim of this study is to make clear the effect of temperature on the potentials of silver-silver chloride(Ag-AgCl) electrodes in potassium chloride(KCl) and sodium chloride(NaCl) solutions for micro biosensor applications. Non-isothermal coefficients, or thermal coefficient of Ag-AgCl electrodes in the concentrations of 0.1 to 1.0[mol/kg] were experimentally measured in the range of 15 to 45[°C]. The electrode was prepared by electrolytical process in KCl or NaCl solutions. The length and diameter of the electrode are 20[mm] and 3[mm], respectively. The coefficients in KCl solutions were 0.42[mV/°C] in 0.1[mol/kg] and 0.26[mV/°C] in 1.0[mol/kg] and almost the same in NaCl solutions. Isothermal coefficients were also measured. Although the coefficients must theoretically be zero by definition for all temperatures, the difference of potentials between a pair of electrodes which is called as offset voltage appears practically. They were 3∼4[μV/°C] in 0.1[mol/kg] and 0.3[μV/°C] in 1[mol/kg] at the steady state of temperature. The values were depending on the degree of the uniformity of electrode surface.

著者関連情報
© 電気学会 2007