透析患者の血液を対象とした開放終端同軸プローブによる複素比誘電率の測定と透析効果
武田 明高田 和之長尾 尋智王 建青藤原 修
著者情報
ジャーナル フリー

2011 年 131 巻 12 号 p. 2046-2050

詳細
抄録

Before evaluating the quality of hemodialysis from the limited volume of human blood using a commercially available open-ended coaxial probe, we previously measured the complex relative permittivity of pure water from 200 MHz to 6 GHz with respect to its measured liquid volume, and revealed that 1.9 ml water in a beaker with a diameter of 24 mm and a depth of 2 mm gives a variation within ±0.5 % for the real part and ±7 % for the imaginary part. Based on the above finding, we measured the dielectric properties of 2.5 ml whole blood at 25°C for 10 normal healthy subjects and 9 hemodialysis patients. The measured results on healthy subjects show good agreement with the data reported by Gabriel for human blood at 37°C, while they provide different dispersion characteristics of straight lines for their Cole-Cole plots. The measured results on the patients give further different dispersion characteristics in comparison with the healthy subjects. In order to investigate the above differences statistically, the student t-test was conducted to reveal that permittivity at infinite frequency for the Cole-Cole plots is significantly different with a level of 1 % among its averaged values for normal healthy subjects and patients before dialysis.

著者関連情報