<生体医工学・福祉工学>
口腔咽喉音分析に基づく無呼吸・低呼吸の無拘束モニタリング
浦南 翔一酒井 徳昭大石 利光松村 雅史
著者情報
ジャーナル フリー

2016 年 136 巻 7 号 p. 1011-1012

詳細
抄録
In this study, a breath monitoring system is being developed using a contact type microphone attached the neck and measuring the sound at the time of respiration (breath sound). In this paper, I made s sound measuring amplifier with adjustable frequency bandwidth amplification degree, so it fits each subject properly as well as improving the sound measuring circuit. Then I demonstrated that normal breathing, apnea and hypopnea could be determined analysing the relation between nasal airflow and the amplitude of the sound produced at neck level doing respiration. In addition, I verified the efficacy of the breath monitoring system by comparing it to an equipment that can, in a simple manner, examine the sleep apnea.