完全埋め込み型糖応答性蛍光ゲルセンサの研究に関するレビュー
高橋 正幸許 允禎興津 輝竹内 昌治
著者情報
ジャーナル フリー

2012 年 132 巻 12 号 p. 437-442

詳細
抄録

This review covers our studies about fully implantable sensor using glucose responsive fluorescence hydrogel based on our previous publications. We succeeded to fabricate the fluorescence hydrogel micro-beads and micro-fibers that have capability to be implanted to the ear of a mouse. Implanted fluorescence hydrogel sensors tracked the fluctuation of blood glucose concentration in the in vivo tests, and the glucose traceability has maintained until 140 days from the implantation with micro-fibers. PEG-bonded micro-fibers successfully suppressed the inflammation. The fluorescence measurement system has the great potential to realize the minimally-invasive and long-term implantable continuous glucose monitoring.