電気学会論文誌E(センサ・マイクロマシン部門誌)
Online ISSN : 1347-5525
Print ISSN : 1341-8939
ISSN-L : 1341-8939
論文
形状記憶合金ワイヤーを用いた低侵襲医療用一方向能動屈曲機構とワイヤーの電気抵抗変化を用いたフィードバック制御
小林 拓生松永 忠雄芳賀 洋一
著者情報
ジャーナル フリー

2015 年 135 巻 10 号 p. 393-398

詳細
抄録
The bending mechanism of a conventional endoscope operates by drawing wires from outside the human body. The shaft structure of endoscopes is relatively complex to avoid buckling caused by wire traction. In this study, one directional bending mechanisms using shape memory alloy (SMA) wires have been developed for minimally invasive medicine. A large bending angle was accomplished by positioning a working point of driven SMA wires near a center point of the mechanism and by surrounding the SMA wires and an inner tube by helical thin metal wire. External diameter and length of the bending part of the mechanism are 4.1 mm and 35 mm respectively. The curvature radius of the fabricated device is 20 mm while maintaining the surface temperature of the device under 41°C in environmental temperature (38°C). Servo control has been considered to be suitable for the application of active bending mechanisms as micro robots. Feedback control by changing in the electrical resistance of the SMA wire was utilized for precise and safty actuation.
著者関連情報
© 2015 電気学会
前の記事 次の記事
feedback