電気学会論文誌E(センサ・マイクロマシン部門誌)
Online ISSN : 1347-5525
Print ISSN : 1341-8939
ISSN-L : 1341-8939
特集論文
折畳式Finger-Poweredマイクロ流体デバイスを用いた唾液中尿酸センシングシステム
工藤 寛之柴原 卓哉東條 良紀
著者情報
ジャーナル 認証あり

2022 年 142 巻 3 号 p. 38-42

詳細
抄録

Here, we report a novel biosensing system using a finger-powered microfluidic device, which is suitable for daily self-testing of uric acid in saliva. To realize the rapid determination of salivary uric acid, the biosensing system measures directly from the paper-based saliva sampling device. The microfluidic device has the diaphragm pump, the reagent reservoir, and the reaction cell in which the electrochemical uric acid biosensor is embedded. Once the microfluidic chip has been folded, the finger-powered microfluidic device prepares the reagent containing the saliva, which is ready to measure. A wireless potentiostat for electrochemical measurement of uric acid with the biosensor was also developed. Combining these elements together, the whole process from saliva sampling to determination was completed in less than a minute (Sensitivity: 0.03 nA/µM). The measurement results were also consistent with that measured using a commercially available uric acid determination kit. Hence, our method is expected to have the potential to enhance the use of biochemical information in the field of healthcare IoT.

著者関連情報
© 2022 電気学会
前の記事 次の記事
feedback
Top