Japanese Journal of Medical Technology
Online ISSN : 2188-5346
Print ISSN : 0915-8669
ISSN-L : 0915-8669
Technical Articles
Evaluation of ACTH assay based on chemiluminescence enzyme immunoassay and comparison of its reactivity with high-molecular-weight ACTH
Yukari MIYAZAWATakashi HERAIRiichi KAWASAKIShigeyuki ENDOYukako KOIKEAkiko YONEYAMA
Author information
Keywords: AIA-CL2400, ACTH, Pro-ACTH, CLEIA
JOURNAL FREE ACCESS FULL-TEXT HTML

2019 Volume 68 Issue 3 Pages 507-513

Details
Abstract

Tosoh Corporation developed a novel reagent to be used with AIA-CL automated immunoassay analyzers to quantify adrenocorticotropic hormone (ACTH) on the basis of the principles of chemiluminescence enzyme immunoassay. The coefficients of variation (CVs) for within-run and between-run reproducibilities ranged from 1.1 to 4.4%. They showed good linearity from 0 to 1,790 pg/mL, and the limit of quantitation was 1.650 pg/mL. With its high sensitivity, the assay may be useful for detecting the presence of ACTH-producing tumors that secrete high-molecular-weight ACTH. The measurement time of this method is shorter than that of the existing method, and it is considered highly useful for routine examinations.

Content from these authors
© 2019 Japanese Association of Medical Technologists
Previous article Next article
feedback
Top