Variable Classifications of Glycemic Index Determined by Glucose Meters
Meng-Hsueh Amanda LinMing-Chang WuJenshinn Lin
著者情報
ジャーナル フリー 早期公開

論文ID: 10-08

この記事には本公開記事があります。
詳細
抄録
The study evaluated and compared the differences of glucose responses, incremental area under curve (IAUC), glycemic index (GI) and the classification of GI values between measured by biochemical analyzer (Fuji automatic biochemistry analyzer (FAA)) and three glucose meters: Accue Chek Advantage (AGM), BREEZE 2 (BGM), and Optimum Xceed (OGM). Ten healthy subjects were recruited for the study. The results showed OGM yield highest postprandial glucose responses of 119.6 ± 1.5, followed by FAA, 118.4 ± 1.2, BGM, 117.4 ± 1.4 and AGM, 112.6 ± 1.3 mg/dl respectively. FAA reached highest mean IAUC of 4156 ± 208 mg × min/dl, followed by OGM (3835 ± 270 mg × min/dl), BGM (3730 ± 241 mg × min/dl) and AGM (3394 ± 253 mg × min/dl). Among four methods, OGM produced highest mean GI value than FAA (87 ± 5) than FAA, followed by BGM and AGM (77 ± 1, 68 ± 4 and 63 ± 5, p<0.05). The results suggested that the AGM, BGM and OGM are more variable methods to determine IAUC, GI and rank GI value of food than FAA. The present result does not necessarily apply to other glucose meters. The performance of glucose meter to determine GI value of food should be evaluated and calibrated before use.