中国・四国整形外科学会雑誌
Online ISSN : 1347-5606
Print ISSN : 0915-2695
ISSN-L : 0915-2695
原著
上腕骨近位部骨折に対するLocking Humeral Spoon Plateの使用経験と問題点
塩田 直史衣笠 清人道中 泰典上田 英輝鄭 明守吉鷹 輝仁松尾 真嗣
著者情報
ジャーナル 認証あり

2004 年 16 巻 2 号 p. 175-180

詳細
抄録
Thirty-six patients (mean age, 64 years) with proximal humeral fractures were treated using Locking Humeral Spoon Plate (LHSP)(Synthes, Japan) from 2001 to 2003. The purpose of this study was to investigate the clinical results of this method. Using the AO/ASIF classification, 6 fractures were classified as 11-A2, 10 were A3, 12 were B1, 4 were B2, 1 was B3, 2 were C1 and 1 was C3.
The average postoperative flexion was 121.7°, the average abduction was 111.4°and the average external rotation was 44.8°. Thirty three patients healed without complication. One patient healed with varus deformity over 5°. In two patients, the plate cut out necessitated revision surgeries in which the LHSP was replaced with a longer plate, or humeral head prosthesis.
The LHSP maintained angular stability between plate and screws, and allowed exercise immediately after the operation. It is important to position the plate to avoid the medial side of the intertubercular sulcus, and to avoid impingement of the shoulder motion. It is possible to stabilize fractures of the 11-C3 type with the LHSP.
The LHSP internal fixation system offered us a new approache for treatment of these proximal humeral fractures.
著者関連情報
© 2004 中国・四国整形外科学会
前の記事 次の記事
feedback
Top