イレウスチューブを用いた小腸運動計測システムの応答性能と解析手法の初期評価
細川 慎二長縄 明大関 健史岡 潔眞部 紀明春間 賢
著者情報
ジャーナル オープンアクセス

2025 年 13 巻 2 号 p. 110-121

詳細
抄録
No functional testing methods to observe the recovery process of patients undergoing treatment with an ileus tube have yet been proposed. In this study, we developed a system that measures small intestine motility as changes in balloon pressure attached to an ileus tube during treatment, using a pressure sensor. As a preliminary technical evaluation of the system, we conducted measurements on five healthy volunteers: three received metoclopramide to promote motility, and two received butylscopolamine bromide to suppress it. We assessed whether the system, along with its signal analysis methods—specifically, time-domain and time–frequency analysis using wavelet transform—could quantitatively detect drug induced changes in motility. After metoclopramide administration, pressure fluctuations increased by approximately 0.2–0.4 kPa in the 1–2 cycles/min range, along with increased spectral power and activity. Conversely, approximately 27 seconds after butylscopolamine bromide administration, a pressure drop of 0.5–1.0 kPa and decreased activity in the 1–2 and 6–11 cycles/min ranges were observed. These results suggest that the system, combined with appropriate signal analysis methods, is technically capable of quantifying motility changes and may serve as a useful tool for clinical monitoring.