運動負荷中の心電図とSpO2 脈波を用いた非侵襲連続心拍出量測定のエコードップラーに対する一致度とトレンドの追従性
須郷 義広酒井 智之寺尾 真美鵜川 貞二落合 亮一
著者情報
ジャーナル フリー

2013 年 83 巻 3 号 p. 273-282

詳細
抄録
A new technology called estimated continuous cardiac output (esCCO) uses pulse wave transit time (PWTT) obtained from an electrocardiogram and pulse oximeter to measure cardiac output (CO) noninvasively and continuously. This study was performed to evaluate the accuracy of esCCO during exercise testing. We compared esCCO with CO measured by echo Doppler aortic velocity-time integral (VTIao_CO). The correlation coefficient between esCCO and VTIao_CO was r = 0.87 (n = 72). Bias and precision were 0.33 ± 0.95 L/min with a percentage error of 30.2%. And the correlation coefficient between change in esCCO and change in VTIao_CO was r = 0.75 (n = 70). The esCCO could detect change in VTIao_CO larger than 1 L/min with a concordance rate of 88.2%. The areas under ROC curves to detect VTIao_CO changes more than 15% and 30% were 0.81 and 0.91, respectively. Polar plots show that the mean polar angle was 2.2° and radial limits of agreement were ±43.6°. The trending ability of esCCO is comparable with that of currently available arterial waveform analysis methods. In this study, the validity of esCCO during exercise testing was assessed and shown to be acceptable with respect to agreement and trending ability compared with the echo Doppler.
著者関連情報
© 2013 日本医療機器学会