日本医学放射線学会物理部会誌
Online ISSN : 2186-9847
Print ISSN : 0288-5506
ISSN-L : 0288-5506
CTスキャナのスライス厚の測定-被写体の精密な移動による
今村 恵子楠本 伸弥藤井 正道宇治 輝幸石川 徹
著者情報
ジャーナル フリー

1984 年 4 巻 2 号 p. 99-105

詳細
抄録
New method is proposed for the measurement of the CT slice thickness. This method is to measure the CT number of a traveling material, placed perpendicularly to the slice plane, as a function of the traveling distance. Movement of the rod was controlled using a traveling microscope with an accuracy of 0.5 mm, The rod was inserted from the outside to the inside of the scan plane. CT number of the imaged rod increases gradually, initially from the air CT number, finally to the genuine CT number of the rod. Full slice width is evaluated as the difference of moving distances. As the rod is projected in a parallel way with the detector aperture, the partial volume effect due to the aperture size is no problem. This method provides a direct and accurate estimate of slice thickness. Comparative study has been made with the 45 degree aluminium strip method.
著者関連情報