医学物理
Online ISSN : 2186-9634
Print ISSN : 1345-5354
ISSN-L : 1345-5354
次世代PET装置におけるデータ収集システムの基礎的検討
吉田 英治村山 秀雄清水 啓司北村 圭司
著者情報
ジャーナル フリー

2003 年 23 巻 1 号 p. 65-72

詳細
抄録
We are under development of a 3D PET scanner with depth of interaction (DOI) capable of high sensitivity and high resolution. In this paper, we estimated the data transfer rate of data acquisition system of this PET scanner by the simulation studies. A maximum data transfer rate of coincidence pair's event information is 10 Mcps and one coincidence event is a 64 bit data format. A data acquisition system, which fulfills the specification of this scanner, is considered for parallel collection with banks including several coincidence units. Simulator of parallel collection system composed of two stand alone simulators and the server PC. Each stand alone simulator is composed of data generation PC, data processing circuit, SCSI and data acquisition PC. As the result of simulation studies, the maximum data transfer rate of the stand alone simulator was 12.0 MB/s, and the maximum data transfer rates at the server PC with one and two nodes were 11.0 and 21.8 MB/s, respectively. Data transfer rate was improved by parameter optimization of a message size and multi processing of data acquisition software using RAM DISK.
著者関連情報
© 一般社団法人 日本医学物理学会
前の記事 次の記事
feedback
Top