ポジトロンカメラによるRIビームの体内到達深度の測定に有効なポジトロン放出核種についての検討
水野 秀之井関 康浦壁 恵理子須田 充金沢 光隆北川 敦志富谷 武浩中村 譲金井 達明石井 慶造
著者情報
ジャーナル フリー

2004 年 24 巻 2 号 p. 37-48

詳細
抄録
The 1°C and 11C beam stop position in a homogeneous phantom was measured using the range verification system in HIMAC. This system was developed to clear uncertainty of beam range within the patient body in heavy ion radiotherapy. In this system, a target is irradiated with RI beams (11C or 10C) and the distribution of the beam end-lpoints are measured by a positron camera. To inspect the precision of the measurement, three experiments were done, simple PMMA phantom irradiation, empirical beam stop position measurements using a range shifter and boundary irradiation using PMMA and lung phantom. Results of the first two experiments were consistent. Consequently, a 0.2 mm standard deviation of statistical error measurement was possible with 250 determinations. For the third experiment, we compared the precision using 10C and 11C beams. The boundary of the PMMA and lung phantom was irradiated with both beams to maximize the positron range effect in the beam range measurement. Consequently, no significant difference was observed between the two beams in spite of the different positron range. Thus, we conclude that the 10C beam was useful for clinical application because of its good statistics owing to the short half-life.