An ambulatory bruxism recording system with sleep-stage analyzing function
Takahiro MizumoriShinji InanoMasakazu SumiyaYasuyoshi KobayashiTakao WatamotoHirofumi Yatani
著者情報
ジャーナル オープンアクセス

2009 年 53 巻 3 号 p. 150-154

詳細
抄録
Purpose: The aim of this study was to develop an ambulatory bruxism recording system capable of sleep-stage analysis.
Methods: A portable EMG system was used to record masseter muscle activity. An EMG sensor was attached onto the masseter muscle belly at either side. EMG data were stored on a notebook type personal computer. A sound level meter was used to assess the sound level of bruxism. Sound level (dB) readings were taken every second and recorded on the same computer. A prototype of sleep sensor, a wristwatch-style biological signal sensor-recorder device, recorded and stored pulse wave, acceleration and temperature on a memory card. All stored data were transferred to a personal computer and analyzed.
Results: The whole system was transportable within a protective case and weighed approximately 5 kg. Raw EMG signals were processed to derive integrated EMG data. TOSHIBA Sleep Analysis Program classified sleep-stages as awake, shallow sleep, deep sleep and REM based on the activity of the autonomic nervous system that was estimated from the fluctuations of pulse intervals. An EMG, sound level and sleep-stage analysis program was developed to analyze all data simultaneously. Using this program, the masseter muscle activity, sound level and sleep-stage could be quantified and correlated.
Conclusion: We developed an ambulatory bruxism recording system that analyzes sleep-stage. We expect that this system will enable us to measure sleep bruxism activity in each sleep-stage on an electromyographical and auditory basis at the subject's home.
著者関連情報

この記事は最新の被引用情報を取得できません。

© 2009 Japan Prosthodontic Society.

This is an open-access article distributed under the terms of Creative Commons Attribution License 4.0 (CCBY 4.0), which allows users to distribute and copy the material in any format so long as attribution is given to the Japan Prosthodontic Society.
https://creativecommons.org/licenses/by/4.0/